7-19
IM 760301-01E
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
Trend Display
Explanation
• ON/OFF
A cursor can be placed on the displayed trend to measure and display the value at
any point. It can be used to measure the data on the vertical axis (Y-axis) and
horizontal axis (X-axis) of various sections of the trend.
• ON: Performs cursor measurement.
• OFF: Does not perform cursor measurement.
• Item under Measurement
Select the target trend for the cursor measurement from T1 to T16.
• Moving the Cursor
• Cursors move on the selected trend.
• Specify the number of points from the left end of the screen to move the cursor with
the left and right ends of the screen assumed to be point 0 and 1001, respectively.
• The cursors moves along the data points displayed on the screen one point at a
time.
• Measurement Item
• Y+: Vertical value of (Y-axis value)
• Yx: Vertical value of cursor x (Y-axis value)
•
∆
Y: The difference between the Y-axis values of and cursor x
• X+: X-axis value of from the left end of the screen
Indicates the relative time of from the left end of the screen with the
left end of the screen taken to be 0 s.
• Xx: X-axis value of cursor x from the left end of the screen
Indicates the relative time of cursor x from the left end of the screen with the
left end taken to be 0 s.
•
∆
X: The difference between the X-axis values of and cursor x
• D+: Date and time at the position
Indicates the date and time (year/month/date hour:minute:second) of
measurement.
• Dx: Date and time at the cursor x position
Indicates the date and time (year/month/date hour:minute:second) of
measurement.
Note
•
if immeasurable data exists, *** is displayed in the measured value display area.
•
∆
Y is measured even if the unit is different. The result has no unit in this case.