2-22
IM 760301-01E
• Normal Integration Mode
The integration time is set in relative time. The integration is stopped after the
specified time elapses or when the integration value reaches the maximum or
minimum integration display value. The integration time and value are held at that
point.
Integration
time
Reset
Start
Integration
value
Hold
Hold
Timer value
• Repetitive Integration Mode (Continuous Integration)
The integration time is set in relative time. When the specified time elapses, the
operation is automatically reset and restarted. Integration is repeated until STOP is
pressed. When the integration value reaches the maximum or minimum integration
display value, the integration time and value are held.
Start
Reset
Integration
time
Integration
value
Hold
Hold
Timer
value
Timer
value
Timer
value
Press STOP
2.6 Integration