2-15
IM 760301-01E
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
Faunctional Description
2.4
Power Measurement
The measured data of voltage, current, power, and so on can be displayed by setting the
screen to numeric display. By dividing the screen into top and bottom halves, waveform
(section 2.7), trend (section 2.8), bar graph
*1
(section 2.8), waveform computation
*2
(section 2.9), and FFT
*2
(section 2.9) can be displayed simultaneously.
*1 Only on models with the advanced computation (/G6) or harmonic measurement (/G5) option.
*2 Only on models with the advanced computation (/G6) option.
Display Resolution
The display resolution for voltage, current, active power, apparent power, reactive power,
and so on is 600000. When the range rating (rated value of the specified range) is
applied, the
Σ
function of voltage, current, active power, apparent power, reactive power,
and so on is set to the decimal point position and unit of the element with the lowest
display resolution of the target elements. For the display resolution during integration,
see section 5.13.
Numeric Display Related to Power Measurement «For procedures, see section 5.1.»
• Selecting the Number of Displayed Items
The number of items can be selected from 4, 8, 16, All, Single List,
*3
and Dual List.
*3
*3 Only on models with the advanced computation (/G6) or harmonic measurement (/G5) option.
• Example in Which Eight Items Are Displayed
Measurement function
Data
• Example in Which All Items Are Displayed
Element and wiring system
Measurement function
Data