5-40
IM 760301-01E
Explanation
For functional details, see section 2.6.
To perform integration, the integration mode and integration time must be set before
starting the operation.
• Selecting the Normal Integration Mode or Repetitive Integration Mode
In this mode, the integration time is set in relative time and the integration is
performed over the specified time. This mode consists of the following two types.
• Normal Integration Mode
The integration time is set in relative time. The integration is stopped after the
specified time elapses, when the maximum integration time of 10000 hours is
exceeded, when STOP is pressed, or when the integration value reaches the
maximum or minimum integration display value (see section 5.11). The integration
time and value are held at that point.
• Repetitive Integration Mode (Continuous Integration)
The integration time is set in relative time. When the specified time elapses, the
operation is automatically reset and restarted. Integration is repeated until STOP is
pressed. When the integration value reaches the maximum or minimum integration
display value, the integration time and value are held.
• Setting the Integration Timer
• The time in units of hour:minute:second can be set in the following range.
00000:00:00 to 10000:00:00
Note
If the integration timer is set to 00000:00:00 in normal integration mode, the integration is
performed in manual integration mode (see section 2.6 and 5.14).
For a description of starting, stopping, and resetting the integration and other items, see
section 5.11.
Note
Integration cannot be started, stopped, or reset in measurement modes other than normal
measurement mode.
5.13 Setting the Normal Integration and Repetitive Integration