Faunctional Description
2.2
Measurement Modes and Measurement
Functions
Measurement Modes
The normal measurement mode on the WT3000 allows you to measure voltage, current,
power, etc. If the advanced computation (/G6) option is installed, measurements can be
performed in a mode specific to each of the functions listed below in addition to the
normal measurement mode. Switch the measurement mode according to your
application.
• Normal Measurement Mode (Normal Mode)
This mode is used to measure voltage, current, power, and integrated values. If the /
G6 option is installed, harmonic measurement
*1
as well as bar graph and vector
displays become available in normal measurement mode.
*1 Equivalent to the /G5 (harmonic measurement) option. This harmonic measurement does not
conform to IEC61000-3-2.
• Wide Bandwidth Harmonic Measurement Mode (Wide-Band Harmonics Mode) (/
G6 Option)
Harmonic measurement up to 50
th
order can be performed on a signal with a 1-kHz
fundamental frequency. Use this mode for harmonic measurement of signals whose
fundamental frequency is higher than the commercial power supply frequency.
• IEC Harmonic Measurement Mode (IEC Harmonics Mode) (/G6 Option)
This mode in combination with the Harmonic/Flicker Measurement Software (761922)
sold separately allows you to perform harmonic measurement conforming to
IEC61000-3-2. Use this mode to check that the harmonics of electric home appliances
and office automation equipment comply with the IEC standards.
• Waveform Computation Mode (Math Mode) (/G6 Option)
This mode allows you to perform various computations on the waveform sampling
data of the input signal. For example, the waveform sampling data of the voltage and
current input signals can be multiplied together to display the waveform of
instantaneous power.
• FFT Mode (/G6 Option)
This mode allows the power spectrum of the input signal to be displayed through FFT
(fast Fourier transform). Use this mode to check the frequency distribution of the input
signal.
• Voltage Fluctuation and Flicker Measurement Mode (Flicker Mode) (/FL Option)
This mode allows you to perform voltage fluctuation and flicker measurements
conforming to IEC61000-3-3. You can use the Harmonic/Flicker Measurement
Software for the WT3000 (761922) to execute measurements or set measurement
conditions from a PC and create a report of measurement results.
• Cycle-by-Cycle Measurement Mode (/CC Option)
This mode allows the voltage, current, power, and other parameters for each cycle of
the AC input signal to be measured.