2-20
IM 760301-01E
Phase Difference «For procedures, see section 5.10»
The display format of the phase difference between the voltage and current of each
element can be selected. With the voltage of each element as a reference, one format
displays the phase difference using 360
°
in the clockwise direction, and the other format
displays lead up to 180
°
in the counterclockwise direction (D) and lag up to 180
°
in the
clockwise direction (G). The phase difference of harmonic order 1 to 100 of the voltage
and current on the harmonic measurement option is displayed using a 360
°
format or a
180
°
format (negative sign for lead and no sign for lag).
Delta Computation «See the
Expansion Function User’s Manual IM760301-51E.»
Delta computation can be performed on models with the delta computation function
option.
Equation for Distortion Factor «See the
Expansion Function User’s Manual IM760301-
51E.»
The distortion factor can be computed on models with the advanced computation (/G6)
or harmonic measurement (/G5) option. You can select one of two equations available
for computing the distortion factor.
2.5 Computation