For functional details, see section 2.6.
To perform integration, the integration mode and integration time must be set before
starting the operation.
• Selecting the Real-time Normal Integration Mode or Real-time Repetitive
Integration Mode
In this mode, integration is performed between the times specified by integration start
and stop. This mode consists of the following two types.
• Real-Time Normal Integration Mode
The date, time, and integration timer time of start and stop of the integration
operation is set. The integration is stopped at the specified time, when the
maximum integration time of 10000 hours is exceeded, or when the integration
value reaches the maximum or minimum integration display value (see section
5.11). The integration time and value are held at that point.
• Real-Time Repetitive Integration Mode (Continuous Integration)
The start and stop of the integration operation is set through date and time and the
integration timer time. The integration is repeated at the specified timer setting
during that time. When the time specified by the timer elapses, the operation is
automatically reset and restarted. The integration is stopped at the specified time
or when the integration value reaches the maximum or minimum integration display
value. The integration time and value are held at that point.
• Setting the Integration Timer
• The time in units of hour:minute:second can be set in the following range.
00000:00:00 to 10000:00:00
Note
If you set the integration timer time to 00000:00:00 in real-time normal integration mode, the
integration starts at the specified start date/time. The integration is stopped at the specified
time, when the maximum integration time of 10000 hours is exceeded, or when the
integration value reaches the maximum or minimum integration display value. The integration
time and value are held at that point.
• Setting the Reservation Time
Set the time to start/stop the integration in year:month:day, hour:minute:second. Make
sure the reservation time for stopping the integration is after the reservation time for
starting the integration.
The selectable range of the values is as follows:
Year:
4-digit year
Hour:minute:second: 00:00:00 to 23:59:59
Note
•
The reservation time allows up to 31 days to be specified for February. In this case, an error
message is displayed at the time of execution of the integration operation (section 5.13).
Reset the reservation time.
•
Leap years are recognized at the time of execution of the integration operation.
For a description of starting, stopping, and resetting the integration and other items, see
section 5.11.
5.14 Setting the Real-Time Integration and Real-Time Repetitive Integration