11.2 Error Messages and Corrective Actions
Error Messages
There are cases in which error messages are displayed on the screen during operation.
This section describes the meanings of the messages and their corrective actions. The
messages can be displayed either in English or Japanese (see section 3.18). If the
corrective action requires servicing, contact your nearest YOKOGAWA dealer for repairs.
In addition to the error messages listed below, there are communication error messages.
These messages are described in the
Communication Interface User’s Manual
(IM760301-17E)
.
Error in Execution
Code
Message
Corrective Action
Reference Section
11
Cannot measure PLL frequency.
Check input level.
–
*
601
Invalid file name.
Check the file name.
9.3, 9.4
602,
No USB device or no storage media inserted. Check the USB device connection, and the
9.1
603
existence of a storage medium in the device.
604
Media failure.
Check the storage medium.
–
605
File not found.
Check the file name and the storage medium.
–
606
Media is protected.
Set the disk’s (medium’s) write protect switch to
–
OFF.
607
Media was removed while accessing.
Check the storage medium.
–
608,
File already exists.
–
9.3
609
610
Contains invalid characters.
–
9.3
611,
Media full.
Delete unnecessary file(s) or use another disk.
9.2 to 9.7
612
613
Cannot delete a directory if there are files
Delete all the files in the directory to be deleted.
9.7
in the directory.
614
File is protected.
–
9.6
615
Physical format error.
Reformat the medium. If the same error occurs,
9.2
the instrument is probably unable to execute a
format on this medium.
616
File system failure.
Check using another disk. If the same message
–
to 620
still appears, maintenance service is required.
621
File is damaged.
Check the file.
–
622
File system failure.
Check using another disk. If the same message
–
to 641
still appears, maintenance service is required.
643
Media failure.
Check the medium.
–
to 653
655
File system failure.
Check using another disk. If the same message
–
to 664
still appears, maintenance service is required.
665
Cannot load this file format. File was
–
–
stored on other models or other versions.
666
File is now being accessed.
Execute after access is made.
–
671
Save data not found.
Check for presence of data.
9.3, 9.4
674
Initializing file system.
Please wait.
9.2
675
Cannot load this file.
–
–
Model/options do not conform.
679
Printer error.
Maintenance service is required.
–
680
Close the cover of the printer unit.
–
–
*
681
Paper empty.
Load a roll chart.
–
*
*
See the
Communication Interface User’s Manual IM760301-17E
.