MedWeld 4000 Overview
1 - 10
Modified: 12/16/08
MedWeld 4000 Operator’s Guide
M-032116 Programs #T03300, #T02300 and #T02400
Timer Assembly
The timer assembly (WTC p/n 900-822x-2) provides a firing signal to
the firing card, as required by the weld schedule. The firing card then
provides the gate signals to the SCRs. The SCRs send welding voltage
to the weld transformer, to apply weld current to the work piece.
The timer assembly houses these major components:
•
Local I/O and
•
An optional FieldBus daughter board.
The following illustrates the connectors on the timer assembly board:
Figure 4:
Timer Assembly Board
Connectors
Normal / Program I/O jumper
MedLAN
port
Analog 2
channel
Analog 1
channel
Power and Local
I/O connector
Connector to
Firing Card
Connectors
to Discrete
J2
J4
J5
J8
J6
J7
J9
J10
(#1)
(#2)
900-822x-2
I/O Boards
5-position
Phoenix connector
A B
C D
E
MedLAN communications LED
Status LEDs (behind J10):
A:
B:
C:
D:
E:
Shunt Trip output
Isolation contactor output
Isolation contactor status
Thermal switch
Control Stop