C-Link Protocol Commands
Datalogging
Thermo Fisher Scientific
Model 48
i
Instruction Manual
B-21
erec
er
xy
x
= | 0 | 1 |
: Reply termination format (see “set format
format
”
command)
y = | 0 | 1 | 2 | : Output format (see “set erec format
format
” command)
These commands return a snapshot of the main operating conditions
(measurements and status) at the time the command is issued. The
following example shows a typical response.
The format is defined within the command (in the case of er
xy
) or by the
current settings of the "format" and "erec format" commands (in the case
of erec). For details on erec formatting, see the “
Record Layout Definition
”
section at the end of this appendix. For details on how to decode the flag
fields within these records, see the "flags" command.
Send:
erec
Receive: erec
09:48 04-06-05 flags 9C040510 CO 0.000 4 loCO -0.002
4 s/r 0.000 los/r 0.902 biasv -112.668 agci 96.500 intt
34.023 cht 47.995 smpfl 0.000 pres 0.000 avgt1 10 avgt2
10 CObkg -0.000 COcoef 1.000 loCOcoef 1.000 COrange
10000000.000 loCOrange 10000000.000 motor 100.000
lrec
srec
lrec
rec num
srec
rec num
lr
xy rec num
sr
xy rec num
lrec
aa:bb oo-pp-qq yy
srec
aa:bb oo-pp-qq yy
rec
= The starting record index number (1=most recent)
num
= The number of records to return (1 to 10)
x
= | 0 | 1 |
: Reply termination format (see “set format
format
”
command)
y = | 0 | 1 | 2 | : Output format (see “set lrec/srec format
format
”
command)
aa
= hours (01 to 23)
bb
= minutes (01 to 59)
oo
= month (01 to 12)
pp
= day (01 to 31)
= year