Operation
Range Menu
3-14
Model 48
i
Instruction Manual
Thermo Fisher Scientific
The CO Range screen defines the concentration range of the analog
outputs. For example, a CO range of 0–50 ppm restricts the analog output
to concentrations between 0 and 50 ppm.
The display shows the current CO range. The next line of the display is
used to change the range. The range screen is similar for the single, dual,
and auto range modes. The only difference between the screens are the
words “High” or “Low” displayed to indicate which range is displayed. The
following example shows the CO range screen in single mode. For more
information about the dual and auto range modes, see “
Single Range
Mode
”, “
Dual Range Mode
”, and “
Auto Range Mode
” earlier in this
chapter.
Table 3–5
lists the available preset ranges.
●
In the Main Menu, choose Range >
Range
.
RANGE:
CURRENTLY: 10000
SET TO: 5000 ?
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
Table 3–5.
Available Operating Ranges
ppm
mg/m
3
1 1
2 2
5 5
10 10
20 20
50 50
100 100
200 200
500 500
1000 1000
2000 2000
5000 5000
CO Range