Operation
Calibration Menu
3-22
Model 48
i
Instruction Manual
Thermo Fisher Scientific
NEXT DATE AND TIME:
19 MAR 2009 12:34 ?
SETTING: DAYS
SET MONTHS
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
CALIBRATE CO:
CO: 15.80
SPAN CONC: 0025.00
0
?
MOVE CURSOR
CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Zero/Span Check menu is available only if the zero/span valve option
is installed. It is used to program the instrument to perform fully
automated zero and span checks or adjustments.
●
In the Main Menu, choose Calibration >
Zero/Span Check
.
ZERO/SPAN CHECK:
>NEXT TIME 01Jan09 12:00
PERIOD HR 24
TOTAL DURATION HR 1.5
ZERO DURATION MIN 30
SPAN DURATION MIN 30
RANGE AVG DIAGS ALARM
PURGE DURATION MIN 30
ZERO/SPAN AVG SEC 60
ZERO CAL RESET ON
SPAN CAL RESET OFF
ZERO/SPAN RATIO 1
The Next Time screen is used to view and set the initial date and time (24-
hour format) of the zero/span check. Once the initial zero/span check is
performed, the date and time of the next zero/span check is calculated and
displayed.
●
In the Main Menu, choose Calibration > Zero/Span Check >
Next
Time
.
Zero/Span Check
Next Time
NEXT DATE AND TIME:
19 MAR 2009 12:34
PRESS TO EDIT
RANGE AVG DIAGS ALARM