Operation
Service Menu
Thermo Fisher Scientific
Model 48
i
Instruction Manual
3-81
The Flow Calibration submenu is used to calibrate the flow sensor to zero,
span, or restore factory default values. The flow calibration screen is visible
only when the instrument is in service mode. For more information on the
service mode, see “
Service Mode
” earlier in the chapter.
Note
This adjustment should only be performed by an instrument service
technician.
▲
●
In the Main Menu, choose Service >
Flow Calibration
.
FLOW SENSOR CAL:
>ZERO 200
SPAN 1.0000
SET DEFAULTS
RANGE AVG DIAGS ALARM
The Calibrate Flow Zero screen calibrates the flow sensor at zero flow.
Note
The pump must be disconnected before performing the zero
calibration.
▲
Note
Wait at least 30 seconds for the reading to stabilize before saving the
value.
▲
●
In the Main Menu, choose Service > Flow Calibration >
Zero
.
CALIBRATE FLOW ZERO:
CURRENTLY: 0.000 L/min
SET TO: 0.000 L/min ?
DISCONNECT PUMP AND
SAVE CURRENT FLOW
RANGE AVG DIAGS ALARM
The Calibrate Flow Span screen allows the user to view and set the flow
sensor calibration span point.
Flow Calibration
Calibrate Flow Zero
Calibrate Flow Span