Operation
Alarms Menu
3-70
Model 48
i
Instruction Manual
Thermo Fisher Scientific
INTERNAL TEMPERATURE:
ACTUAL MIN: 8.0
o
C
SET MIN TO: 10.0
o
C ?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Bench Temperature screen displays the current bench temperature and
sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 40 to 59 °C. If the bench temperature reading goes beyond
either the minimum or maximum limit, an alarm is activated and the alarm
(bell) icon appears in the status bar on the Run screen and in the Main
Menu.
●
In the Main Menu, choose Alarms >
Bench Temp
.
BENCH TEMPERATURE:
ACTUAL 48.4
o
C
>MIN 40.0
o
C
MAX 59.0
o
C
RANGE AVG DIAGS ALARM
The Minimum Bench Temperature alarm limit screen is used to change the
minimum bench temperature alarm limit. The minimum and maximum
bench temperature screens function the same way.
●
In the Main Menu, choose Alarms > Bench Temp >
Min
or
Max
.
BENCH TEMPERATURE:
ACTUAL MIN: 40.0
o
C
SET MIN TO: 44.0
o
C ?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Pressure screen displays the current reaction chamber pressure reading
and sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 250 to 1,000 mmHg. If the pressure reading goes beyond either
Bench Temperature
Min and Max Bench
Temperature Limits
Pressure