Operation
Instrument Controls Menu
3-54
Model 48
i
Instruction Manual
Thermo Fisher Scientific
Concentrations
Corrected Concentrations
Other Measurements
Analog Inputs
O
2
Sensor Temp
The Analog Input Configuration menu displays a list of the 8 analog input
channels available for configuration. This screen is only displayed if the I/O
expansion board option is installed. Configuration includes entering the
Descriptor, Units, Decimal Places, choice of 1-10 points in the table, and
corresponding number of points selected.
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config
.
ANALOG INPUT CONFIG:
>CHANNEL 1 IN1
CHANNEL 2 IN2
CHANNEL 3 IN3
CHANNEL 4 IN4
CHANNEL 5 IN5
RANGE AVG DIAGS ALARM
ANALOG INPUT 01 CONFIG:
>DESCRIPTOR IN1
UNITS V
DECIMAL PLACES 2
TABLE POINTS 2
POINT 1
RANGE AVG DIAGS ALARM
POINT 2
The Descriptor screen allows the user to enter the descriptor, or name, for
the selected analog input channel. The descriptor is used in datalogging
and streaming data to report what data is being sent out. The descriptor
may be from 1 to 3 characters in length, and defaults to IN1 to IN8 (user
input channel number).
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > select Channel >
Descriptor
.
Analog Input
Configuration
Descriptor