Operation
Instrument Controls Menu
Thermo Fisher Scientific
Model 48
i
Instruction Manual
3-55
ANALOG INPUT DESCRIPTOR:
CURRENTLY: IN1
I
N1
ABCDEFGHIJKLMN BKSP
OPQRSTUVWXYZ PAGE
0123456789 ./- SAVE
RANGE AVG DIAGS ALARM
The Units screen allows the user to enter the units for the selected analog
input channel. The units are displayed on the diagnostic screen and in
datalogging and streaming data. The units may be from 1 to 3 characters in
length, and defaults to V (volts).
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > select Channel >
Units
.
ANALOG INPUT UNITS:
CURRENTLY: V
V
N1
ABCDEFGHIJKLMN BKSP
OPQRSTUVWXYZ PAGE
0123456789 ./- SAVE
RANGE AVG DIAGS ALARM
The Decimal Places screen allows the user to select how many digits are
displayed to the right of the decimal, from 0 to 6, with a default of 2.
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > Select Channel >
Decimal Places
.
DECIMAL PLACES:
CURRENTLY: 2
SET TO: 3 ?
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
The Number of Table Points screen allows the user to select how many
points are used in the analog input conversion table for the selected
channel. The instrument uses linear interpolation between the points in
this table to determine what the reading value is based on the analog input
voltage. Each point in the table consists of an analog input voltage value
(0–10.5 V) and a corresponding reading value. Only two points are
Units
Decimal Places
Number of Table Points