Operation
Instrument Controls Menu
Thermo Fisher Scientific
Model 48
i
Instruction Manual
3-29
ERASE LREC LOG FILE DATA?
ERASE
ARE YOU SURE YOU WANT TO?
x
PRESS
TO CONFIRM ERASURE
RANGE AVG DIAGS ALARM
The Erase Log screen is used to erase all saved data for the selected record
type only (not both srecs and lrecs).
●
In the Main Menu, choose Instrument Controls > Datalogging Settings
>
Erase Log
.
The Select Content submenu displays a list of 32 record fields to use to
configure the data to be logged. When a field is selected, a submenu list
appears of the items to choose from to assign to those fields. Item types are
Concentrations, Corrected Concentrations (if the O
2
compensation option
is enabled), Other Measurements, and Analog Inputs (if the I/O expansion
board is installed). These record fields comprise a temporary list of items
for the selected record type that must be committed via the datalogging
menu before the changes will apply. Note that committing any changes to
this list will erase all currently logged data for the selected record type only,
as the format of the stored data is changed.
●
In the Main Menu, choose Instrument Controls > Datalogging Settings
>
Select Content
.
LREC FIELDS:
>FIELD 1 NO
FIELD 2 CO
FIELD 3 PRES
FIELD 4 PMTT
FIELD 5 INTT
RANGE AVG DIAGS ALARM
The Choose Field Data submenu displays a list of the types of data that can
be logged for the current field. Choices are Concentrations, Corrected
Concentrations (if the O
2
compensation option is enabled), Other
Measurements, and Analog Inputs (if the I/O expansion board is installed).
Erase Log
Select Content
Choose Field Data
ERASE LREC LOG FILE DATA?
ERASE
RANGE AVG DIAGS ALARM