Operation
Service Menu
Thermo Fisher Scientific
Model 48
i
Instruction Manual
3-83
INITIAL S/R RATIO:
CURRENTLY: 1.161111
MEASURED: 1.161111
SET TO: 1.16211
0
?
MOVE CURSOR
CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
Up to three gas concentrations (cal-points) for each range may be calibrated
to using the following steps. Three cal-points will give the most accurate
readings over the entire range. The calibration process is sequential and will
work properly if all steps are followed in order. The following example
shows the multi-point calibration screen in single range mode. In dual or
auto range modes, “HI” or “LO” is displayed to indicate the calibration of
the high or low concentrations and function the same way.
Note
This adjustment should only be performed by an instrument service
technician.
▲
●
In the Main Menu, choose Service >
Multipoint Calibration
.
MULTI-POINT CAL:
>CALIBRATE POINT 1
CALIBRATE POINT 2
CALIBRATE POINT 3
COEFFICIENTS
DEFAULT COEFFICIENTS
RANGE AVG DIAGS ALARM
The Calibrate Point 1 screen allows the user to view and set the selected
calibration point. The calibrate 2 and calibrate 3 screens function the same
way.
●
In the Main Menu, choose Service > Multipoint Cal >
Calibrate 1, 2
or
3
.
Multi-Point
Calibration
Calibrate Point 1/2/3