C-Link Protocol Commands
Alarms
B-14
Model 48
i
Instruction Manual
Thermo Fisher Scientific
alarm agc intensity min
alarm agc intensity max
These commands report the agc intensity alarm minimum and maximum
value current settings. The following example reports that the agc intensity
alarm minimum value is 20.
Send:
alarm agc intensity min
Receive:
alarm agc intensity min 20
set alarm agc intensity min
value
set alarm agc intensity max
value
These commands set the agc intensity alarm minimum and maximum
values to
value
, where
value
is a floating-point number representing agc
intensity alarm limits. The following example sets the agc intensity alarm
maximum value to 20.
Send:
set alarm agc intensity max 20
Receive:
set alarm agc intensity max 20 ok
alarm bias voltage min
alarm bias voltage max
These commands report the bias voltage alarm minimum and maximum
value current settings. The following example reports that the bias voltage
alarm minimum value is -128.
Send:
alarm bias voltage min
Receive:
alarm bias voltage min -128
set alarm bias voltage min
value
set alarm bias voltage max
value
These commands set the bias voltage alarm minimum and maximum
values to
value
, where
value
is a floating-point number representing bias
voltage alarm limits. The following example sets the bias voltage alarm
maximum value to -110.
Send:
set alarm bias voltage max -110
Receive:
set alarm bias voltage max -110 ok
alarm chamber temp min
alarm chamber temp max
These commands report the bench temperature alarm minimum and
maximum value current settings. The following example reports that the
bench temperature alarm minimum value is 35.0 °C.
Send:
alarm chamber temp min
Receive:
alarm chamber temp min 35.0 deg C
Alarms