C-Link Protocol Commands
Diagnostics
B-18
Model 48
i
Instruction Manual
Thermo Fisher Scientific
Table B–4.
Alarm Trigger Values
Value
Alarm Trigger
00 Floor
01 Ceiling
alarm trig conc o2
This command reports the O
2
concentration alarm trigger action for
minimum alarm, current setting, to either floor or ceiling. The following
example shows the O
2
concentration minimum alarm trigger set to ceiling,
according to
Table B–4
.
Send:
alarm
trig
conc
o2
Receive:
alarm trig conc o2 1
set alarm trig conc o2
value
This command sets the O
2
concentration alarm minimum
value,
where
value
is set to either floor or ceiling, according to
Table B–4
. The following
example sets the O
2
concentration minimum alarm trigger to ceiling.
Send:
set alarm trig conc o2 1
Receive:
set alarm trig conc o2 1 ok
agc int
This command reports the current intensity of the reference channel AGC
circuit. The following example reports that the current AGC intensity is
200,000 Hz.
Send:
agc
int
Receive:
agc int 200000 Hz
bias voltage
This command reports the current IR bias supply voltage. The following
example reports that the bias voltage is -102.3 volts
Send:
bias
voltage
Receive:
bias voltage -102.3 V
diag volt mb
This command reports the diagnostic voltage measurements on the
motherboard. The sequence of voltages is: Positive 24, positive 15, positive
5, positive 3.3, and negative 3.3. Each voltage value is separated by a space.
Send:
diag
volt
mb
Receive:
diag volt mb 24.1 14.9 4.9 3.2 -3.2