Operation
Alarms Menu
Thermo Fisher Scientific
Model 48
i
Instruction Manual
3-77
O2 CONCENTRATION:
ACTUAL MIN: 25.00
%
SET MIN TO: 10.00
%
?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Minimum Trigger screen allows the user to view and set the O
2
concentration alarm trigger type to either floor or ceiling. The minimum
alarm may be programmed as a floor trigger (alarm is triggered when the
concentration falls below the minimum value) or a ceiling trigger (alarm is
triggered when the concentration goes above the minimum value).
●
In the Main Menu, choose Alarms > O2 Concentration >
Min
Trigger
.
MIN TRIG(CEILING/FLOOR:
ACTUAL TRIGGER: CEILING
SET TRIGGER TO: FLOOR ?
TOGGLE AND SAVE VALUE
RANGE AVG DIAGS ALARM
The External Alarms menu is used to display the external alarm status for
alarm 1, 2, and 3 only if the external alarm is assigne as a digital input.
External alarms can be assigned to digital inputs using the Instruments
Controls > I/O Configuration > Digital Input Settings menu. External
alarms can also be logged, streamed, or output as a digital output by
choosing “EXT ALARMS” from the Other Measurements list.
●
In the Main Menu, choose Alarms >
External Alarms
.
ALARMS:
EXTERNAL ALARM 1 OK
EXTERNAL ALARM 2 OK
EXTERNAL ALARM 3 OK
RANGE AVG DIAGS ALARM
Min Trigger
External Alarms