Operation
Service Menu
Thermo Fisher Scientific
Model 48
i
Instruction Manual
3-89
O2 CORRECTION CONC:
CURRENTLY: 15.55 %
SET TO: 16.00 % ?
MOVE CURSOR
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Dilution Ratio screen allows the user to view and set the dilution ratio.
Acceptable values are 1–500: 1. The default is 1:1. When this value is set,
the dilution ratio is applied to all concentration measurements. This screen
is only accessible if the dilution ratio option is installed and when the
instrument is in service mode. For more information on the service mode,
see “
Service Mode
” earlier in the chapter.
●
In the Main Menu, choose Service >
Dilution Ratio
.
DILUTION RATIO:
CURRENTLY: 001.00 :1
SET TO: 00
2
.00 :1 ?
MOVE CURSOR
CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Setup Run Screens menu is used to define and edit up to five custom
Run screens. When the instrument is turned on, screen 1 (or zero) is the
default screen. The Setup Run Screens menu is visible only when the
instrument is in the Service mode. For more information on the service
mode, see “
Service Mode
” earlier in the chapter.
●
In the Main Menu, choose Service >
Setup Run Screens
.
SELECT RUN SCREEN:
>RUN SCREEN 1
RUN SCREEN 2
RUN SCREEN 3
RUN SCREEN 4
RUN SCREEN 5
RANGE AVG DIAGS ALARM
START SCREEN
DEFAULT
Dilution Ratio
Setup Run Screens