Operation
Alarms Menu
3-68
Model 48
i
Instruction Manual
Thermo Fisher Scientific
Note
If the analyzer is in service mode, pressing
will toggle the
selected item between yes and no (with the exception of dilution ratio and
auto calibration, which may be only enabled at the factory).
▲
●
In the Main Menu, choose Diagnostics >
Instrument Configuration
.
INSTRUMENT CONFIGURATION:
>I/O EXPANSION BOARD YES
ZERO/SPAN VALVES YES
ZERO AIR SCRUBBER YES
WHEEL PURGE YES
DILUTION NO
RANGE AVG DIAGS ALARM
AUTO CALIBRATION NO
O2 SENSOR YES
The Contact Information screen displays the customer service information.
●
In the Main Menu, choose Diagnostics >
Contact Information
.
CONTACT INFORMATION:
CALL CENTER: 508-520-0430
WEB: WWW.THERMO.COM
RANGE AVG DIAGS ALARM
The Alarms menu displays a list of items that are monitored by the
analyzer. If the item being monitored goes outside the lower or upper limit,
the status of that item will go from “OK” to either “LOW” or “HIGH”,
respectively. If the alarm is not a level alarm, the status will go from “OK”
to “FAIL”. The number of alarms detected is displayed to indicate how
many alarms have occurred. If no alarms are detected, the number zero is
displayed.
To see the actual reading of an item and its minimum and maximum
limits, move the cursor to the item and press
.
Items displayed are determined by the options installed. The zero/span
check and auto calibration screens are visible only if the zero/span check or
auto calibration options are enabled. The O
2
Concentration Alarm menu
item is only displayed if the internal O
2
sensor option is installed. The
motherboard status, interface board status, and I/O Expansion board status
Contact Information
Alarms Menu