Servicing
Flow Transducer Calibration
Thermo Fisher Scientific
Model 48
i
Instruction Manual
7-25
4.
Remove the two retaining screws holding the flow transducer to the
floor plate and remove the flow transducer (
Figure 7–12
).
Figure 7–12.
Replacing the Flow Transducer
5.
Install the new flow transducer following the previous steps in reverse.
6.
Calibrate the flow transducer. Refer to the “
Flow Transducer
Calibration
” procedure that follows.
Use the following procedure to calibrate the flow transducer.
Equipment Required:
Calibrated flow sensor
WARNING
The service procedures in this manual are restricted to qualified
service representatives.
▲
If the equipment is operated in a manner not specified by the
manufacturer, the protection provided by the equipment may be
impaired.
▲
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Remove the instrument cover.
2.
Disconnect the pump cable from AC PUMP connector on the
measurement interface board.
Flow Transducer
Calibration
Retaining Screws (2)