Servicing
Filter Wheel Replacement
7-12
Model 48
i
Instruction Manual
Thermo Fisher Scientific
Brass Connectors (2)
Cover Screws (2)
Clamp Screws (2)
Figure 7–5.
Replacing the IR Source
Use the following procedure to replace the filter wheel.
Equipment Required:
Filter wheel
Allen wrenches, 5/32-inch and 5/64-inch
Philips screwdriver
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
Filter Wheel
Replacement
SEMS Nuts (4)
Washers (2)
IR Source Cover
IR Source Mounts
IR Source
Motor Plate
Access Hole
Filter Wheel