C-Link Protocol Commands
Calibration
B-28
Model 48
i
Instruction Manual
Thermo Fisher Scientific
Send:
CO
coef
Receive:
CO coef 1.200
set co coef
value
set high co coef
value
set low co coef
value
These commands set the CO coefficients to a user-defined
value
, where
value
is a floating-point representation of the coefficient. The following
example sets the CO coefficient to 1.200.
Send:
set CO coef 1.200
Receive:
set CO coef 1.200 ok
coef o2
This command reports the O
2
coefficient. The example below reports that
the O
2
coefficient is 1.000.
Send:
coef
o2
Receive:
coef o2 1.000
set coef o2
value
This command sets the O
2
coefficient to a user-defined
value
, where
value
is a floating-point representation of the coefficient. The example below sets
the O
2
coefficient to 1.005.
Send:
set coef o2 1.005
Receive:
set coef o2 1.005 ok
coef 0
coef 1
coef 2
high coef 0
high coef 1
high coef 2
low coef 0
low coef 1
low coef 2
The “coef 0”, coef 1”, and coef 2” commands report the coefficients of the
curve developed from the Hi Multi-Point Calibration in single range mode.
The “high coef 0”, high coef 1”, and high coef 2” commands report the
coefficients of the curve developed from the Hi Multi-Point Calibration in
dual or auto range mode. The “low coef 0”, low coef 1”, and low coef 2”
commands report the coefficients of the curve developed from the Lo
Multi-Point Calibration in dual and auto range mode. The following
example reports the coefficient 0 having a value of 1.005.
Send:
coef
0
Receive:
coef 0 1.005 ok