In the dual range mode, there are two independent analog outputs. These
are labeled simply as the “High Range” and the “Low Range”. Each
channel has its own analog output range, averaging time, and span
coefficient.
This enables the sample concentration reading to be sent to the analog
outputs at two different ranges. For example, the low CO analog output
can be set to output concentrations from 0 to 50 ppm and the high CO
analog output set to output concentrations from 0 to 100 ppm.
In addition, each CO analog output has a span coefficient. There are two
span coefficients so that each range can be calibrated separately. This is
necessary if the two ranges are not close to one another. For example, the
low CO range is set to 0–50 ppm and the high CO range is set to 0–10000
ppm.
By default, in the dual range mode, the analog outputs are arranged on the
rear panel connector as shown in
Figure 3–5
. See
Table 3–3
for channels
and pin connections. Dual range mode may be selected from the “
Range
Mode Select
” on page
3-78
.