Operation
Password
Thermo Fisher Scientific
Model 48
i
Instruction Manual
3-95
The Set Password screen is used to set the password that is used to unlock
the front panel, and is shown if the instrument is unlocked and the
password is not set.
●
In the Main Menu, choose Password >
Set Password
ENTER NEW PASSWORD:
R
ANGXXXXXE
A
BCDEFGHIJKLMN BKSP
OPQRSTUVWXYZ PAGE
0123456789 ./- SAVE
RANGE AVG DIAGS ALARM
The Lock Instrument screen is used to lock the instrument’s front panel so
users cannot change any settings from the front panel. The lock instrument
screen is shown if the instrument is unlocked and the password is set.
●
In the Main Menu, choose Password >
Lock Instrument
LOCK FRONT PANEL:
XXX
PRESSING ENTER WILL
XXXX
PREVENT USER FROM CHANGING
X
CONFIG FROM FRONT PANEL
XX
LOCK AND RETURN TO RUN
RANGE AVG DIAGS ALARM
If the instrument keyboard is locked via the front panel using Password >
Lock Instrument
, the instrument reports being in Remote mode. In this
mode, the keypad is locked, data can be viewed but not changed using the
front panel interface, and the remote “Set” commands are active.
If the instrument keyboard is unlocked via the front panel using Password
>
Unlock Instrument
, the instrument reports being in Local mode, the
front panel interface is unlocked, and data can be changed from the front
panel.
Refer to “
C-Link Protocol Commands
” appendix for detailed information
about “mode”, “allow mode”, and “power up mode” commands.
Set Password
Lock Instrument
Lock/Unlock and
Local/Remote Operation