Servicing
Bench Heater Assembly Replacement
Thermo Fisher Scientific
Model 48
i
Instruction Manual
7-17
5.
Calibrate the instrument. Refer to the “
Calibration
” chapter in this
manual.
Figure 7–8.
Replacing the Optical Switch (Inverted View)
Use the following procedure to replace the bench heater assembly (
Figure
7–8
).
Equipment Required:
Bench heater
Heat conductive compound
Flatblade screwdriver, 1/4-inch
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover.
2.
Remove the optical bench following the “
Optical Bench Replacement
”
procedure in this chapter.
Bench Heater
Assembly
Replacement
Heater Board Assy
Optical Switch
Heater
Heater