Servicing
Capillary Cleaning or Replacement
7-26
Model 48
i
Instruction Manual
Thermo Fisher Scientific
3.
From the Main Menu, choose Service >
Flow
Calibration
.
The Flow Sensor Cal screen appears.
Note
If Service Mode is not displayed, refer to “
Accessing the Service
Mode
” on page
7-4
, then return to the beginning of this step.
▲
4.
At the Flow Sensor Cal screen, press
to select
Zero
.
The Calibrate Flow Zero screen appears.
5.
Wait at least 10 seconds for the zero reading to stabilize, and then press
to save the zero flow value.
6.
Reconnect the pump cable to the AC PUMP connector on the
measurement interface board.
7.
Connect a calibrated flow sensor at the SAMPLE bulkhead on the rear
panel.
8.
Press
to return to the Flow Sensor Cal screen.
9.
At the Flow Sensor Cal screen, press
to select
Span
.
The Calibrate Flow Span screen appears.
10.
Wait at least 10 seconds for the reading to stabilize, use
and
to enter the flow sensor reading, and press
to
save the value.
11.
Install the cover.
Use the following procedure to clean or replace the capillary (
Figure 7–13
).
Equipment Required:
Wrench, 5/8-inch
Capillary
Capillary cleaning wire (smaller than 0.015-inch)
Capillary Cleaning
or Replacement