Operation
Diagnostics Menu
Thermo Fisher Scientific
Model 48
i
Instruction Manual
3-63
INTERFACE BOARD VOLTAGES:
> 3.3 SUPPLY 3.3 V
5.0 SUPPLY 5.0 V
15.0 SUPPLY 15.0 V
24.0 SUPPLY 24.1 V
-15.0 SUPPLY -15.0 V
RANGE AVG DIAGS ALARM
18.0 IR SUPPLY 18.0 V
18.0 MOT SUPPLY 18.0 V
BIAS SUPPLY -110.0 V
The I/O Board screen (read only) is used to display the current voltage
readings on the I/O expansion board. This menu is only displayed if the
I/O expansion board option is installed.
●
In the Main Menu, choose Diagnostics > Voltages >
I/O Board
Voltages
.
I/O BOARD VOLTAGES:
3.3 SUPPLY 3.3 V
5.0 SUPPLY 5.0 V
24.0 SUPPLY 24.0 V
-3.3 SUPPLY -3.3 V
RANGE AVG DIAGS ALARM
The Temperatures screen (read only) displays the current internal
instrument temperature and bench temperature. The internal temperature
is the air temperature measured by a sensor located on the interface board.
●
In the Main Menu, choose Diagnostics >
Temperatures
.
TEMPERATURES:
INTERNAL 34.6
o
C
BENCH 49.7
o
C
RANGE AVG DIAGS ALARM
The Pressure screen (read only) displays the current optical bench pressure.
The pressure is measured by a pressure transducer.
●
In the Main Menu, choose Diagnostics >
Pressure
.
I/O Board Voltages
Temperatures
Pressure