Servicing
Fuse Replacement
Thermo Fisher Scientific
Model 48
i
Instruction Manual
7-9
3.
Remove the cover.
4.
Disconnect the three connectors that pass through the center of the
partition panel that connect to the measurement interface board.
5.
Remove two screws from the left side rear of the case (viewed from
front).
6.
Remove one screw from the bottom rear of the case.
7.
Remove one screw from the top front of the partition panel.
8.
While holding the case securely, loosen the captive screw at the rear of
the measurement case assembly, and pull the measurement case
assembly from the rear of the case.
9.
Remove the screw at the top rear of the partition panel that secures the
top of the panel to the measurement case assembly, and lower the
panel, being careful not to put excessive tension on the cables.
10.
Re-install the measurement case assembly by following the previous
steps in reverse.
Use the following procedure to replace the fuses.
Equipment Required:
Replacement fuses:
250VAC, 3 Amp, SlowBlow (for 100VAC and 110VAC models)
250VAC, 1.60 Amp, SlowBlow (for 220-240VAC models)
1.
Turn the instrument OFF and unplug the power cord.
2.
Remove fuse drawer, located on the AC power connector.
3.
If either fuse is blown, replace both fuses.
4.
Insert fuse drawer and reconnect power cord.
Fuse Replacement