Operation
Service Menu
3-84
Model 48
i
Instruction Manual
Thermo Fisher Scientific
SET CAL GAS POINT:
RANGE: 10000
CAL FACTOR 1: 1.0000
CAL GAS 1: 0040.00
0
?
MOVE CURSOR
CHANGE VALUE SAVE
RANGE AVG DIAGS ALARM
The Coefficients screen allows the user to view and re-calculate the
calibration coefficients.
●
In the Main Menu, choose Service > Multipoint Cal > select Cal Point
>
Coefficients
.
COEFFICIENTS:
1: 1.00000
2: 1.00000
3: 1.00000
RECALCULATE ?
RANGE AVG DIAGS ALARM
The Default Coefficients screen allows the user to view and reset the
calibration coefficients to default values.
●
In the Main Menu, choose Service > Multipoint Cal > select Cal Point
>
Default Coefficients
.
COEFFICIENTS:
1: 1.00000
2: 1.00000
3: 1.00000
SET TO DEFAULTS ?
RANGE AVG DIAGS ALARM
The Pre-amp Board Calibration screen is used to adjust the preamp board
calibration parameters. The preamp board calibration screen is visible only
when the instrument is in service mode. For more information on the
service mode, see “
Service Mode
” earlier in this chapter.
Note
This adjustment should only be performed by an instrument service
technician.
▲
Coefficients
Default Coefficients
Preamp Board
Calibration