Operation
Diagnostics Menu
Thermo Fisher Scientific
Model 48
i
Instruction Manual
3-61
Note
The current timezone may say NULL before the timezone is set for
the first time, or if the timezone was cleared with a C-Link command.
▲
●
In the Main Menu, choose Instrument Controls >
Timezone
.
TIMEZONE FOR TIMESERVER:
CURRENTLY: UTC (GMT)
SET TO: EST (GMT+5) ?
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
The Diagnostics menu provides access to diagnostic information and
functions. This menu is useful when troubleshooting the instrument. The
analog input readings and analog input voltages are only displayed if the
I/O expansion board option is installed. The O
2
Sensor Readings screen is
only displayed if the internal O
2
sensor option is installed.
●
In the Main Menu, choose
Diagnostics
.
DIAGNOSTICS:
>PROGRAM VERSIONS
VOLTAGES
TEMPERATURES
PRESSURE
FLOW
RANGE AVG DIAGS ALARM
SAMPLE/REF RATIO
AGC INTENSITY
MOTOR SPEED
O2 SENSOR READINGS
ANALOG INPUT READINGS
ANALOG INPUT VOLTAGES
DIGITAL INPUTS
RELAY STATES
TEST ANALOG OUTPUTS
INSTRUMENT CONFIGURATION
CONTACT INFORMATION
The Program Versions screen (read only) shows the version number of the
programs installed. Prior to contacting the factory with any questions
regarding the instrument, please note the product model name and the
program version numbers.
●
In the Main Menu, choose Diagnostics >
Program Versions
.
Diagnostics
Menu
Program Versions