Operation
Diagnostics Menu
3-62
Model 48
i
Instruction Manual
Thermo Fisher Scientific
PROGRAM VERSIONS:
PRODUCT: MODEL 48i
VERSION: 01.06.08.248
FIRMWARE: 11.08.105
RANGE AVG DIAGS ALARM
The Voltages menu displays the current diagnostic voltage readings. This
screen enables the power supply to be quickly read for low or fluctuating
voltages without having to use a voltage meter. The I/O board is only
displayed if the I/O expansion board option is installed.
●
In the Main Menu, choose Diagnostics >
Voltages
.
VOLTAGES:
>MOTHERBOARD
INTERFACE BOARD
I/O BOARD
RANGE AVG DIAGS ALARM
The Motherboard screen (read only) is used to display the current voltage
readings on the motherboard.
●
In the Main Menu, choose Diagnostics > Voltages >
Motherboard
Voltages
.
MOTHERBOARD VOLTAGES:
3.3 SUPPLY 3.3 V
5.0 SUPPLY 5.0 V
15.0 SUPPLY 15.0 V
24.0 SUPPLY 24.1 V
-3.3 SUPPLY -3.3 V
RANGE AVG DIAGS ALARM
The Interface Board screen (read only) is used to display the current voltage
readings on the interface board. This menu is only displayed if the I/O
expansion board option is installed.
●
In the Main Menu, choose Diagnostics > Voltages >
Interface Board
Voltages
.
Voltages
Motherboard Voltages
Interface Board Voltages