C-Link Protocol Commands
Alarms
B-16
Model 48
i
Instruction Manual
Thermo Fisher Scientific
Send:
set alarm conc o2 max 100
Receive:
set alarm conc o2 max 100 ok
alarm internal temp min
alarm internal temp max
These commands report the internal temperature alarm minimum and
maximum value current settings. The following example reports that the
internal temperature alarm minimum value is 15.0 °C.
Send:
alarm internal temp min
Receive:
alarm internal temp min 15.0 deg C
set alarm internal temp min
value
set alarm internal temp max
value
These commands set the internal temperature alarm minimum and
maximum values to
value
, where
value
is a floating-point number
representing internal temperature alarm limits in degrees C. The following
example sets the internal temperature alarm maximum value to 45.0 °C.
Send:
set alarm internal temp max 45
Receive:
set alarm internal temp max 45 ok
alarm motor speed min
alarm motor speed max
These commands report the motor speed alarm minimum and maximum
value current settings. The following example reports that the motor speed
alarm minimum value is 99.8 %.
Send:
alarm motor speed min
Receive:
alarm motor speed min 99.8
set alarm motor speed min
value
set alarm motor speed max
value
These commands set the motor speed alarm minimum and maximum
values to
value
, where
value
is a floating-point number representing motor
speed alarm limits in minutes. The following example sets the motor speed
alarm maximum value to 100.2 %.
Send:
set alarm motor speed max 100.2
Receive:
set alarm motor speed max 100.2 ok
alarm pressure min
alarm pressure max
These commands report the pressure alarm minimum and maximum value
current settings. The following example reports that the pressure alarm
minimum value is 650 mmHg.