160
Technical Specifications
Genie_TS_Series GigE Vision Camera
Iris Connector – DC Mode
Pin
Signal
Direction
Definition
1
LENS-Control-
Out
Lens Control – (3.6V – 48mA)
2
LENS-
Out
Lens C
3
LENS-DRV+
Out
Lens Drive + (3.6V – 48mA)
4
LENS-DRV-
Out
Lens Drive -
Input Signals Electrical Specifications
External Inputs Block Diagram
External Input Details
Opto-coupled (2.4V to 24V) with internal current limit.
Selectable input trigger threshold levels for TTL, 12V, and 24V signal inputs (see
feature).
Used as trigger acquisition event, counter or timestamp event, or integration control.
User programmable debounce time from 0 to 255µs in 1µs steps.
Source signal requirements:
Single-ended driver meeting TTL, 12V, or 24V standards
Differential signal drivers cannot be used due to the shared input common
External Input Timing Reference
Input Level
Standard
Maximum Input
Frequency
Minimum Pulse
Width
Source Current
Requirements
Maximum Signal Propagation Delay
at 60°C
Input Signal Direction
0 to 3.3V
17 µs
TTL (3.3V)
86 kHz
9.5 µs
16 mA
3.3V to 0
19 µs
TTL (5.0V)
43 kHz
5.6 µs
16 mA
0 to 12V
10 µs
12V
360 kHz
4.8 µs
16 mA
12V to 0
12 µs
0 to 24V
9 µs
24V
230 kHz
4.1 µs
16 mA
24V to 0
11 µs