Important functions (selection)
7.3 Calibration and validation functions
SIPROCESS UV600
94
Operating Instructions, 10/2013, A5E03529375-02
Menu call:
Main menu → Calibrations → Settings Calib. → Sample gas purging time
The following menu is displayed (example):
Lines 3 and 4 show the minimum and maximum values of the sample gas purging time in
seconds; the duration can be edited in line 2 as described in section .
7.3.2.2
Configuring automatic calibration
This function can be used to 'compose' an automatic calibration procedure.
Menu call:
Main menu → Calibrations → Settings → Automatic
The following menu is displayed with the names of the possible automatic procedures and
their activation status (example).
Use the
and
keys to select the desired procedure, and activate this by pressing
the
key.