Important functions (selection)
7.5 Parameter assignments
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
155
7.5.11.2
Status output for the switchover phase
The logic status of the assigned digital output indicates whether the purging time or the
measuring time is active during the automatic measuring point switchover.
Digital output status
Description
0
Purging time
1
Measuring time
Menu call:
Main menu
→ Parameters → Additional functions → Measuring point switchover → Status
(MPS)
In the menu, use the
,
and
keys to select the desired digital output.
In the follow-up menu, use the
,
and
keys to select the desired switching
logic. The following meanings apply:
●
inverted = inv
●
direct = no display
7.5.11.3
Selecting the measuring point display
Menu call:
Ma
in menu → Parameters → Additional functions → Measuring point switchover → Measured
values
Example of the displayed menu:
Setting the measured component
1.
Use the
,
and
keys to select the desired component.
2.
Use the
,
and
keys to set the desired state.