Important functions (selection)
7.3 Calibration and validation functions
SIPROCESS UV600
106
Operating Instructions, 10/2013, A5E03529375-02
The following menu is displayed (example):
You can now set the desired value as described in section Character input (Page 64).
Setting the purging time for calibrations
The purging time is the waiting time between switching over to the calibration gas and
commencement of the measurement. It can be individually set for each measured
component in each calibration gas setting.
CAUTION
Risk of incorrect calibrations
If the purging time is set too short, this can falsify the calibration results.
•
In case of doubt, you should rather select the purging time to be too long and not too
short.
Menu call:
Main menu → Calibrations → Settings → Calibration gases → Calibration gas setting → Use →
Measured component→ Purging time
The following menu is displayed (example):
Figure 7-2
3511514_Use_NO_Purging_time