Important functions (selection)
7.5 Parameter assignments
SIPROCESS UV600
154
Operating Instructions, 10/2013, A5E03529375-02
Entering the name of the measuring point
1.
Use the
,
and
keys to select 'Name'
2.
Set the name in the follow-up menu as described in section Character input (Page 64).
Setting the purging time
1.
Use the
,
and
keys to select 'Purging time'
2.
Set the purging time in the follow-up menu as described in section Character input
(Page 64). Information on the correct setting of purging times can be found in section
Setting the sample gas purging time (Page 93).
Setting the measuring time
1.
Use the
,
and
keys to select 'Name'
2.
Set the measuring time in the follow-up menu as described in section Character input
(Page 64).
Assigning a status output to the measuring point
The assigned digital output is to control a solenoid valve which switches over the sample gas
path to this measuring point. The digital output is logically active as long as the measuring
point is activated. The electronic switching logic can be inverted.
1.
Use the
,
and
keys to select 'Status'
2.
In the follow-up menu, use the
,
and
keys to select the desired digital
output.
3.
In the follow-up menu, use the
,
and
keys to select the desired switching
logic. The following meanings apply:
–
inverted = inv
–
direct = no display