Important functions (selection)
7.4 Maintenance functions
SIPROCESS UV600
126
Operating Instructions, 10/2013, A5E03529375-02
7.4.5
Restart BCU
Restarting the control unit (BCU) can be meaningful if unclear function faults are present.
Measured value processing then restarts; all saved values remain unchanged.
NOTICE
Risk for connected systems
The normal measuring function is temporarily interrupted during the restart, and the
maintenance signal is activated. The analyzer is ready for operation again following
execution of the restart.
▸ Make sure that this situation does not cause any problems with connected positions.
Menu call:
Main menu → Maintenance → Restart BCU
The following menu is displayed (example):
Press the
key to trigger the restart.