Description
3.5 Materials wetted by sample gas
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
19
3.5
Materials wetted by sample gas
Table 3- 1
Component
Material
Analyzer unit (sample chamber):
Aluminum or stainless steel mat. no. 1.4404 [*], epoxy
resin
Optical window
CaF
2
or quartz [*], epoxy resin
Gas path, gaskets
FKM (Viton™), PTFE, stainless steel mat. no. 1.4571 [*]
Chamber:
Aluminum or stainless steel [*]
Gas inlet/outlet
PVDF, stainless steel mat. no. 1.4401 [*]
Moisture sensor
Stainless steel mat. no. 1.4571, platinum, epoxy resin
Diaphragm pump
Pump body
Diaphragm
PVDF
FKM (Viton™), EPDM
[*] Depending on version