Important functions (selection)
7.3 Calibration and validation functions
SIPROCESS UV600
104
Operating Instructions, 10/2013, A5E03529375-02
The parameters have the following meanings:
Component
Selected component
Active
"Yes" means that this calibration gas configuration is available in other
functions
"No" prevents the use for calibrations and validations. This can be used, for
example, if the real calibration gas is temporarily unavailable or should not be
used.
Concentration Waiting time between activation of calibration gas and commencement of
measurements
As zero gas "Yes" means that this calibration gas can be used for zero calibrations and
validation measurements
As ref. gas
"Yes" means that this calibration gas can be used for full-scale calibrations
and validation measurements
Assignment of a measured component
This setting is used to assign a component to the associated location in the "Use" list of the
calibration gas configuration.
Menu call:
Main menu → Calibrations → Settings → Calibration gases → Calibration gas setting → Use →
Measured component→ Component
The following menu is displayed with the previously set components (real and virtual)
(example). The information on the right-hand side indicates the virtual components for whose
calculation the real components are used. Only one component is displayed if the real and
virtual components are identical.
You can then select the desired component using the
,
and
keys.