Important functions (selection)
7.5 Parameter assignments
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
141
The following menu is displayed (example):
Use the
,
and
keys to set the desired state.
Setting the pump capacity
The setting defines the electric supply to the pump which in turn influences the delivery rate
of the pump. This function is used to set the desired volume flow of sample gas. This is
better than operating the pump at full power and throttling the volume flow using a control
valve. As a result of this measure, there is a reduced load on the pump and the service life is
extended.
Menu call:
Main menu → Parameters → Pump → Pump off
The following menu is displayed (example):
You can now modify the value of the relative pump capacity in line 2 as described in section
Character input (Page 64).