Important functions (selection)
7.5 Parameter assignments
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
153
Holding the measured value
●
If a measuring point is active, the measured values from it correspond to the current value
of the gas analyzer following the purging time.
●
While other measuring points are active, the measured value for this point is displayed as
the constant value which was last measured at this point (hold amplifier function).
Holding of the measured value also functions with analog outputs which output the value
from a measuring point.
7.5.11.1
Programming the measuring point switchover
Menu call:
Main menu → Parameters → Additional functions → Measuring point switchover → Measuring
points (MPi) → Measuring point
The following menu is displayed (example):
The parameters have the following meanings:
On/off
Activation status
Name
Name of measuring point
Purging
time
Set purging time
Measuring
time
Set measuring time
Status
Digital output for status of output range
Activate/deactivate measuring point
1.
Use the
,
and
keys to select 'On/Off'
2.
In the follow-up menu, use the
,
and
keys to set the desired state. The
following meanings apply:
On
This measuring point is used for the measuring point switchover
Off
This measuring point is not used for the measuring point switchover