Important functions (selection)
7.5 Parameter assignments
SIPROCESS UV600
144
Operating Instructions, 10/2013, A5E03529375-02
Configuring an output range
In the first menu, select parameter 3 (Range 0) or 4 (Range 1). You can now:
●
Use the
,
and
keys to activate or deactivate the selected output range:
(on/off).
and
●
use the
,
and
keys to set the selected status
●
Set the start-of-scale value of the output range. To do this:
–
Use the
,
and
keys to select "Start" and
–
set the physical value which is to correspond to the electronic zero.
●
Set the full-scale value of the output range. To do this:
–
Use the
,
and
keys to select "End" and
–
set the physical value which is to correspond to the full-scale value of the electronic
signal span (20 mA).
●
Set the mode of range selection. To do this:
–
Use the
,
and
keys to select the range selection ('Range sel.')
and
–
use the
,
and
keys to set the selected mode.
The following modes are possible:
auto
Autoranging
0
Output range 0 permanently active
1
Output range 1 permanently active
●
Select the digital output for the output range status. To do this:
–
Select "Status"
–
Use the
,
and
keys to select the desired digital output.
–
In the subsequent menu, select the electronic switching logic.
–
In the subsequent menu, use the
,
and
keys to select the electronic
switching logic.