Startup
5.5 Initial calibration
SIPROCESS UV600
52
Operating Instructions, 10/2013, A5E03529375-02
A menu is displayed with the selected calibration or validation function, in this case the
calibration of the full-scale value (reference point).
6.
Select
to start with the calibration.
Table 5- 1
Sequence of a calibration/validation
Manual calibration
Automatic calibration
➝
Wait until the status line displays
"Calibration: Calibration gas" or "Validation: Calibration
gas".
➝ Pass the envisaged calibration gas through the
sample gas inlet
of the SIPROCESS UV600.
➝ Wait until the status line displays
"Calibration: Sample gas" or "Validation: Sample gas".
➝ Connect the sample gas to the
sample gas inlet again.
➝
The automatic procedure starts up.
The currently executed part of the
procedure is displayed in the
status line.
1.
Wait until 'Stop' is displayed in the status line.
This signals the end of the calibration or validation procedure.
2.
To exit the calibration or validation, press the
key.