Important functions (selection)
7.3 Calibration and validation functions
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
107
Lines 3 and 4 show the minimum and maximum values of the purging time in seconds; the
duration can be edited in line 2 as described in section Character input (Page 64).
Using calibration gas as zero gas
These settings define for which calibrations and validations the associated calibration gas is
available as zero gas.
Menu call:
Main menu → Calibrations → Settings → Calibration gases → Calibration gas setting → Use →
Measured component→ As zero gas
The following menu is displayed (example):
You can now set the desired calibration or validation procedure using the
and
keys and move on to the follow-up menu using the
key where you can find information
on the calibration gas availability. The following menu is displayed (example):
You can then set the availability using the
,
and
keys.