Important functions (selection)
7.3 Calibration and validation functions
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
115
Carrying out a validation procedure
CAUTION
Interferences with connected positions
A validation measurement interrupts measuring mode.
•
Inform affected positions prior to a validation of the impending interruption in measuring
mode.
Menu call:
Main menu → Calibrations → Calibration → Measured component → Validation function
The following menu is displayed (example):
Figure 7-3
3211_Val_Zero_point
Proceed as follows:
1.
Connect the calibration gas to the sample gas inlet of the analyzer.
2.
Select
to start with the validation.
The validation procedure commences. The individual validation phases are displayed on the
screen during the procedure. The meanings are as follows:
Phase
Text in the status line
1 - Flow of calibration gas
Purging with cal. gas
2 - Determination of measured value of calibration gas (actual value)
Measuring
3 - Validation: the results are calculated and saved
Calibration: calibration running
Calculating
4 - Flow of sample gas
Purging with sample gas